Sporal Test - Biological Autoclave Check
The sporal test (Geobacillus stearothermophilus) is the biological test that GIS recommends running no less often than once a week. Learn how to run it and record the result.
A sporal test is a biological sterilisation indicator -- the only method that gives a definitive answer to the question: did your autoclave actually kill all microorganisms, including bacterial spores? Chemical indicators that change colour only confirm that a process took place, not that sterilisation was successful. The recommendations of Poland's Chief Sanitary Inspector for the beauty sector (May 2026) call for biological control no less often than once a week, and it is worth recording the result in a sterilisation log. If you are not yet sure how to do this correctly, this article will walk you through the entire process step by step.
What a Sporal Test Is and What the GIS Recommendations Say
A sporal test (biological test, biological sterilisation indicator) is a preparation containing spores of Geobacillus stearothermophilus bacteria. These organisms are exceptionally resistant to high temperature and moisture -- if your autoclave can destroy them, it will destroy all other pathogens as well. That is why the sporal test is the gold standard for verifying sterilisation effectiveness.
The Act of 5 December 2008 on preventing and combating infections and infectious diseases in humans (Article 16) requires you to implement and apply procedures that protect against infection, but it does not set a frequency for biological tests. The frequency comes from the GIS (Chief Sanitary Inspector) recommendations: biological control no less often than once a week. This is a recommendation, not a statutory rule. Write the frequency you choose into your own procedure and follow it. Gaps in documentation can end in a fine, and failing to apply the procedures is itself a petty offence (Article 50(1) of the Act, a fine).
An important distinction: chemical indicators of classes 4, 5 and 6 are process indicators -- they confirm that an instrument was exposed to the appropriate temperature and time, but they do not confirm sterility. The GIS recommendations ask you to assess every cycle with physical, chemical and biological indicators, so one type does not replace the others.
How to Perform a Sporal Test Correctly
There are two types of biological tests available on the market: ampoules sent to an external laboratory (results within 24-72 hours), and self-contained biological indicators (SCBIs), where you carry out the culture yourself using a dedicated incubator after 24 or 48 hours.
Step-by-step procedure:
- Place the biological indicator in the coldest spot of the autoclave chamber -- usually near the drainage outlet, at the lower rear of the chamber. This is the place where steam arrives last and where the temperature is lowest.
- Place the test alongside a normal instrument load -- do not run a separate cycle just for the test.
- Run a full sterilisation cycle according to the parameters for that load. The GIS recommends steam sterilisers that meet at least the PN-EN 13060 standard; choose the device class and cycle programme to match your tools and the autoclave manufacturer's instructions.
- After the cycle ends, remove the indicator and activate it according to the manufacturer's instructions (crushing the ampoule with growth medium in the case of SCBIs).
- Place in the incubator for 24-48 hours at the temperature recommended by the manufacturer (usually 55-60°C), or send it to a laboratory.
- Read the result: no colour change in the medium = negative result = sterilisation successful. Colour change (usually from purple to yellow) = positive result = FAILURE.
The cost of a single biological test ranges from 30 to 80 PLN depending on the manufacturer and method (SCBIs are more expensive but give results faster and without the need to send anything by post).
What to Do with a Positive Result -- the Failure Protocol
A positive biological test result is a serious situation requiring immediate action. It means your autoclave was not sterilising effectively -- at least in that one cycle, but probably for a longer period.
Action protocol:
- Stop using the autoclave immediately. No instrument may leave the device labelled as "sterile" until the fault has been investigated.
- Identify and recall all instruments sterilised since the last negative test. The less often you test, the more instruments are affected, possibly several weeks of work -- all of them must be re-sterilised in another working autoclave or destroyed.
- Contact the autoclave service department or manufacturer. Check: chamber seals, filter, temperature and pressure sensor, safety valve. Request a service report.
- Document the entire incident in the control log: date of failure, cycle number, actions taken, repair date and date of retest.
- Carry out at least 3 further biological tests after the repair to confirm the device is working correctly.
- If you served clients during that time -- you do not need to notify them if instruments were single-use or sterilised in a verified alternative autoclave. If that was not the case, consult a lawyer.
Documentation -- What to Enter in the Log
Every biological test must be recorded in the autoclave control log. This log is an internal salon document that you can present during a Sanepid inspection. The regulations give it neither a name nor a template, as we explain in the article Sterilisation log for a salon: template.
Each entry should include: test date, autoclave serial number, cycle number/identifier, cycle parameters (temperature, time, pressure), type and batch number of the biological indicator, result (negative/positive), and the signature of the responsible person. If the result was positive -- add a description of corrective actions taken and the date of the retest with its result.
The GIS recommendations say records of inspections and maintenance of the equipment should be kept, but they give no number of years, and in the sources we checked we found no rule on how long a salon must keep such a log. Write the period into your own procedure. It is worth photographing indicators (both negative and positive) and archiving the photos together with the log entry -- in case of any dispute.
See also: Sterilisation Log in a Salon: Template.
Frequently Asked Questions (FAQ)
How often should I perform a sporal test?
The GIS recommendations call for biological control no less often than once a week. This is a recommendation, not a statutory rule, so write the frequency into the salon's procedure and keep to it. For testing after a repair or a programme change, follow the autoclave manufacturer's instructions. Earlier versions of this article said "once a month". We corrected that on 3 November 2026 after checking the GIS recommendations.
Do I need to send the test to a laboratory, or can I do it myself?
You can use self-contained biological indicators (SCBIs) with your own incubator -- you read the result yourself after 24-48 hours. You do not need to send anything to a laboratory. It is important that the incubator is calibrated and that you follow the indicator manufacturer's instructions precisely.
What is a class 5 chemical indicator and does it replace a biological test?
It does not replace it. A class 5 indicator (integrator) changes colour when all critical process parameters have been reached (temperature, time, saturated steam). It is an excellent process indicator -- but it indicates the "process", not "sterility". The only proof of sterility is a biological test.
How much does an incubator for biological tests cost?
Incubators for SCBI tests cost from 500 to 2,000 PLN. This is a one-off cost that pays for itself within a few months (you save on laboratory postage). Before buying an incubator, make sure it is compatible with the specific brand of indicators you intend to use.
Can a Sanepid inspection officer take away my autoclave control log?
How an inspector may handle your documents during an inspection is governed by inspection rules that this article does not cover. For safety, keep the log in electronic format or make regular copies of the pages -- in case the original is lost.
Legal status as of: 3 November 2026
Legal basis: Articles 16 and 50(1) of the Act of 5 December 2008 on preventing and combating infections and infectious diseases in humans (consolidated text: Journal of Laws 2025, item 1675; status in the Sejm API: in force; amending acts were not analysed separately); "Hygiene and sanitary recommendations and requirements of the Chief Sanitary Inspector for beauty services" (GIS, May 2026, chapter IV point 2, file from gov.pl downloaded on 3 November 2026). Not checked: the text of standards PN-EN 13060 and PN-EN ISO 17665 (paid standards), cycle parameters for specific autoclaves, rules on inspections and the collection of documents, earlier versions of the GIS recommendations. The Ministry of Health regulations for medical entities do not apply to a salon that performs no medical procedures, so we do not discuss them.
Last reviewed: 3 November 2026
Want Ready-Made Autoclave Control Documentation?
The NailsReady Foundation package (from 249 PLN) includes a ready-made autoclave control log template, a failure protocol template and a maintenance checklist -- to adapt to your autoclave and procedure. You do not have to create documentation from scratch.
See the NailsReady Foundation Package